Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.

Check the device or its packaging for

Model Number: M3-30A. UDI-DI: 00821925005679. All Serial Numbers.
Why
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
Distributed
International distribution to the country of Japan.