TemporaryFDAOngoing

Can cause harm that is usually temporary.

Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No

Check the device or its packaging for

Lot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853
Why
Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.
Distributed
US Nationwide distribution in the states of Colorado and Florida.