Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355.

Check the device or its packaging for

Model Number: BD-420X-1355. UDI-DI: 00821925042261. Lot Number: 460898.
Why
Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,
Distributed
International distribution in the country of Japan.
Company
Olympus Corporation of the Americas
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3216-2026 · retrieved 18 September 2026