Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Check the device or its packaging for

No UDI-DI Lot P22E159
Why
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Distributed
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Company
Howmedica Osteonics Corp.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3195-2026 · retrieved 16 September 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.

Local Recall