Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

Check the device or its packaging for

Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark Digital ECG. The Midmark Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 7817519 7817645, L1130587 V2967564 UDI-DI: 00841709121319, 08056732652491
Why
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
Distributed
Worldwide distribution - US Nationwide and the country of Canada.
Company
Midmark Corporation
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3013-2026 · retrieved 25 August 2026