Can cause harm that is usually temporary.

GETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilator system, for facility use. Maquet Critical Care AB, is a subsidiary of Getinge.

Check the device or its packaging for

DUI/DI 07325710000304, Serial/Lot Numbers: 21049, 21088, 21089, 21093, 21100, 21103 21108, 21114, 21115.
Why
The module did not undergo electrical safety tests and functional tests during production.
Distributed
US Nationwide distribution in the states of CA, GA, MN, NY.