TemporaryFDAOngoing
Can cause harm that is usually temporary.
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
Check the device or its packaging for
UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).
- Why
- Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
- Distributed
- US Nationwide distribution.
- Company
- Tobii Dynavox Llc
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-2960-2026 · retrieved 14 August 2026
Read the official notice