Can cause serious harm or death.

INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.

Check the device or its packaging for

Model/UDI-DI: SC-2317-50/08714729861614, SC-2317-70/08714729861638. All unused products.
Why
Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.
Distributed
US: NY. OUS: Canada, Portugal, Spain, France, Netherlands, Great Britain, Italy, Israel, Serbia, Greece, Germany, United Arab Emirates, Ireland, Hungary, Czech Republic, Brazil, Colombia, Argentina