munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System

Check the device or its packaging for

Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699
Why
A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
Distributed
Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.