Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections

Check the device or its packaging for

Software Version 2.5 UDI-DI code: 15099590802073
Why
Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
Distributed
U.S.: FL, GA, IL, MI, and PA O.U.S.: Belgium, France, Hungary, Poland, Taiwan (Province of China), Tunisia, United Arab Emirates, and United Kingdom of Great Britain and Northern Ireland.
Company
Beckman Coulter Biomedical GmbH
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2834-2026 · retrieved 31 July 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.