Español
TemporaryFDAOngoing

Can make you unwell for a while. Do not eat it.

EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77522

Check your package for

UDI-DI: 00199150023288, Lot Numbers: 0233286436; UDI-DI: 20199150023282, Lot Numbers: 0232709428 0232744924 0232745061 0232745092 0232798556 0232862763 0232862778 0232862794 0232862798 0232903216 0232903291 0232903383 0232990983 0233286318 0233286414 0233286436 0233427120 0233528730 0233553623 0233556386 0233585113 0233585114 0232670802
Why
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Distributed
Worldwide - US Nationwide distribution.
Company
Medtronic Perfusion Systems
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.

Local Recall