Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70

Check the device or its packaging for

UDI-DI: 30695085407007; Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349
Why
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
Distributed
Worldwide - US Nationwide distribution.
Company
ICU Medical Inc.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2759-2026 · retrieved 17 July 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.