Can cause harm that is usually temporary.

Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades

Check the device or its packaging for

FCO72800812; FCO72800814 728305 728306 DI (01)00884838059474 728307 DI (01)00884838104600 728311 728231 (01)00884838083325 728232
Why
The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
Distributed
US Nationwide Distribution including Puerto Rico
Company
PHILIPS MEDICAL SYSTEMS