Español
SeriousFDAOngoing

Can cause serious harm or death.

Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device

Check the device or its packaging for

UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Why
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Distributed
Worldwide distribution - US Nationwide and the country of Canada.
Company
Baxter Healthcare Corporation
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2550-2026 · retrieved 1 July 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.

Local Recall