GraveFDAEn curso
Puede causar una enfermedad grave o la muerte. No lo coma.
Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.
Revise su paquete
UDI: 00887761981492; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
- Motivo
- There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
- Distribuido
- Worldwide distribution - US Nationwide and the country of Canada.
- Empresa
- Baxter Healthcare Corporation
- Anunciado
- El retiro comenzó