LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

Check your package for

Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
Why
Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Distributed
US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.