Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Philips Azurion 5 M12. Model Number: 722227.

Check the device or its packaging for

Model Number: 722227. UDI Numbers: (01)00884838099227(21)135, (01)00884838099227(21)88, (01)00884838099227(21)271, (01)00884838099227(21)127, (01)00884838099227(21)128, (01)00884838099227(21)179, (01)00884838099227(21)270, (01)00884838099227(21)246, (01)00884838099227(21)338, (01)00884838099227(21)108, (01)00884838099227(21)305, (01)00884838099227(21)356, (01)00884838099227(21)346, (01)00884838099227(21)313, (01)00884838099227(21)104, (01)00884838099227(21)109, (01)00884838099227(21)337, (01)00884838099227(21)306, (01)00884838099227(21)215, (01)00884838099227(21)213, (01)00884838099227(21)216, (01)00884838099227(21)226, (01)00884838099227(21)83, (01)00884838099227(21)251, (01)00884838099227(21)268, (01)00884838099227(21)309, (01)00884838099227(21)316, (01)00884838099227(21)290, (01)00884838099227(21)310, (01)00884838099227(21)345, (01)00884838099227(21)116, (01)00884838099227(21)153, (01)00884838099227(21)262, (01)00884838099227(21)204, (01)00884838099227(21)113, (01)00884838099227(21)110, (01)00884838099227(21)256, (01)00884838099227(21)308, (01)00884838099227(21)90, (01)00884838099227(21)200, (01)00884838099227(21)362, (01)00884838099227(21)242, (01)00884838099227(21)136, (01)00884838099227(21)95, (01)00884838099227(21)2, (01)00884838099227(21)221, (01)00884838099227(21)165, (01)00884838099227(21)16, (01)00884838099227(21)12, (01)00884838099227(21)257, (01)00884838099227(21)284, (01)00884838099227(21)248, (01)00884838099227(21)26, (01)00884838099227(21)292, (01)00884838099227(21)58, (01)00884838099227(21)130, (01)00884838099227(21)125, (01)00884838099227(21)184, (01)00884838099227(21)57, (01)00884838099227(21)131. Serial Numbers: 135, 88, 271, 127, 128, 179, 270, 246, 338, 108, 305, 356, 346, 313, 104, 109, 337, 306, 215, 213, 216, 226, 83, 251, 268, 309, 316, 290, 310, 345, 116, 153, 262, 204, 113, 110, 256, 308, 90, 200, 362, 242, 136, 95, 2, 221, 165, 16, 12, 257, 284, 248, 26, 292, 58, 130, 125, 184, 57, 131.
Why
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Distributed
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Company
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-0294-2027 · retrieved 27 October 2026