English
TemporalFDAEn curso

Puede causar daños que por lo general son pasajeros.

Philips Azurion 5 M12. Model Number: 722227.

La agencia publicó este retiro solo en inglés. El nombre del producto y los códigos del paquete se muestran tal como los escribió la agencia, para que coincidan exactamente con su paquete.

Revise el dispositivo o su empaque

Model Number: 722227. UDI Numbers: (01)00884838099227(21)135, (01)00884838099227(21)88, (01)00884838099227(21)271, (01)00884838099227(21)127, (01)00884838099227(21)128, (01)00884838099227(21)179, (01)00884838099227(21)270, (01)00884838099227(21)246, (01)00884838099227(21)338, (01)00884838099227(21)108, (01)00884838099227(21)305, (01)00884838099227(21)356, (01)00884838099227(21)346, (01)00884838099227(21)313, (01)00884838099227(21)104, (01)00884838099227(21)109, (01)00884838099227(21)337, (01)00884838099227(21)306, (01)00884838099227(21)215, (01)00884838099227(21)213, (01)00884838099227(21)216, (01)00884838099227(21)226, (01)00884838099227(21)83, (01)00884838099227(21)251, (01)00884838099227(21)268, (01)00884838099227(21)309, (01)00884838099227(21)316, (01)00884838099227(21)290, (01)00884838099227(21)310, (01)00884838099227(21)345, (01)00884838099227(21)116, (01)00884838099227(21)153, (01)00884838099227(21)262, (01)00884838099227(21)204, (01)00884838099227(21)113, (01)00884838099227(21)110, (01)00884838099227(21)256, (01)00884838099227(21)308, (01)00884838099227(21)90, (01)00884838099227(21)200, (01)00884838099227(21)362, (01)00884838099227(21)242, (01)00884838099227(21)136, (01)00884838099227(21)95, (01)00884838099227(21)2, (01)00884838099227(21)221, (01)00884838099227(21)165, (01)00884838099227(21)16, (01)00884838099227(21)12, (01)00884838099227(21)257, (01)00884838099227(21)284, (01)00884838099227(21)248, (01)00884838099227(21)26, (01)00884838099227(21)292, (01)00884838099227(21)58, (01)00884838099227(21)130, (01)00884838099227(21)125, (01)00884838099227(21)184, (01)00884838099227(21)57, (01)00884838099227(21)131. Serial Numbers: 135, 88, 271, 127, 128, 179, 270, 246, 338, 108, 305, 356, 346, 313, 104, 109, 337, 306, 215, 213, 216, 226, 83, 251, 268, 309, 316, 290, 310, 345, 116, 153, 262, 204, 113, 110, 256, 308, 90, 200, 362, 242, 136, 95, 2, 221, 165, 16, 12, 257, 284, 248, 26, 292, 58, 130, 125, 184, 57, 131.
Motivo
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.La agencia escribió este motivo solo en inglés.La agencia escribió este motivo solo en inglés.
Distribuido
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
Empresa
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Anunciado
Esta es la fecha en que se anunció el retiro. La empresa pudo haberlo iniciado semanas antes.
El retiro comenzó

Fuente: registro Z-0294-2027 de FDA · consultado el 27 de octubre de 2026