Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm

Check the device or its packaging for

UDI: 00817906023992/Lot: GM008380; GM007868
Why
Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
Distributed
US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
Company
In2bones USA, LLC
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-0157-2027 · retrieved 14 October 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.