TemporaryFDAOngoing
Can cause harm that is usually temporary.
CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System
Check the device or its packaging for
UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026
- Why
- Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
- Distributed
- US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
- Company
- Wright Medical Technology, Inc.
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-3334-2026 · retrieved 30 September 2026
Read the official notice