Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Medline Convenience Kits: 1) MAJOR NEURO VCH, Catalog Number: DYNJ39218I; 2) NEURO PACK-LF, Catalog Number: DYNJ41246F; 3) CRANIOTOMY PACK (CRSJE)227-LF, Catalog Number: DYNJ47487A; 4) MAJOR NEURO PACK, Catalog Number: DYNJ57233J; 5) NEURO, Catalog Number: DYNJ82388A; 6) NEURO PACK, Catalog Number: DYNJ86656; 7) NEURO ACD, Catalog Number: DYNJ904549G

Check the device or its packaging for

1) DYNJ39218I, UDI-DI: 10195327013042(each), 40195327013043(case), Lot Number: 22HMA400; 2) DYNJ41246F, UDI-DI: 10195327112011(each), 40195327112012(case), Lot Number: 23GBS783; 3) DYNJ47487A, UDI-DI: 10889942106240(each), 40889942106241(case), Lot Number: 22HMD080; 4) DYNJ57233J, UDI-DI: 10198459723025(each), 40198459723026(case), Lot Number: 26CMG313; 5) DYNJ82388A, UDI-DI: 10195327482749(each), 40195327482740(case), Lot Number: 25KMB622; 6) DYNJ86656, UDI-DI: 10195327663155(each), 40195327663156(case), Lot Number: 24EDB134; 7) DYNJ904549G, UDI-DI: 10193489832686(each), 40193489832687(case), Lot Number: 23IBL172
Why
The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may present health risks that could require medical intervention, such as neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
Distributed
US Nationwide.
Company
Medline Industries, LP
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3268-2026 · retrieved 24 September 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.