Español
SeriousFDAOngoing

Can cause serious harm or death.

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE

Check the device or its packaging for

MDR107000P UDI-DI 00080196288284 All lots
Why
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Distributed
Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
Company
Medline Industries, LP
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3228-2026 · retrieved 30 September 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.