Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0

Check the device or its packaging for

All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)
Why
A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.
Distributed
Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.
Company
Boston Scientific Corporation
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3188-2026 · retrieved 16 September 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.