Español
SeriousFDAOngoing

Can cause serious harm or death.

INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP, Replacement Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-0712, 44-0707, 44-0709, 44-0715.

Check the device or its packaging for

1. Model Number: 44-0712. UDI-DI: (01)60888937009516. Lot Code: 550031624. 2. Model Number: 44-0707. UDI-DI: (01)20888937009488. Lot Code: 550021057. 3. Model Number: 44-0709. UDI-DI: (01)20888937009495. Lot Code: 550031623. 4. Model Number: 44-0715. UDI-DI: (01)20888937009525. Lot Code: 550031625.
Why
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
Distributed
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Company
CooperSurgical, Inc.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3140-2026 · retrieved 18 September 2026