Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C

Check the device or its packaging for

Lots 25JR01801 25KR01225 25KR01251 25NR01016 25NR01265 UDI-DI 00840861102532
Why
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Distributed
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Company
Fresenius Medical Care Renal Therapies Group, LLC
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-3086-2026 · retrieved 2 September 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.