TemporaryFDAOngoing
Can cause harm that is usually temporary.
Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit
Check the device or its packaging for
All lot codes
- Why
- Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
- Distributed
- Worldwide distribution - US Nationwide and the country of Panama.
- Company
- Medline Industries, LP
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began