TemporaryFDAOngoing
Can cause harm that is usually temporary.
Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder
Check the device or its packaging for
Lot Code: Model No 1500-1030; UDI-DI 00840097513225; Lot Numbers A00030, A00065, A09828, A09830 Model No 1500-1035; UDI-DI 00840097513232; Lot Numbers A00064, A09829
- Why
- Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.
- Distributed
- U.S. Nationwide distribution in the state of Texas.
- Company
- restor3d Inc.
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-2987-2026 · retrieved 19 August 2026
Read the official notice