Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131

Check the device or its packaging for

Lot Code: All serial numbers/ GTIN: (01) 00862992000327
Why
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Distributed
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.