Can cause harm that is usually temporary.

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Check the device or its packaging for

1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065
Why
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Distributed
Worldwide distribution - US Nationwide.
Company
Beckman Coulter Inc.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.