Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

Check the device or its packaging for

UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).
Why
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Distributed
US Nationwide distribution.
Company
Tobii Dynavox Llc
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2959-2026 · retrieved 14 August 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.