Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

GEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011 00055445010 00055445011 00055360010.A

Check the device or its packaging for

Part No. 00055360010 UDI-DI 08426950807599 Lots 260505K, 260504AG, 260505F; Part No. 00055360011 UDI-DI 08426950807605 Lots 260505F and 260505J; Part No. 00055407504 UDI-DI 08426950807537 Lot 260505R; Part No. 00055415010 UDI-DI 08426950807698 Lots 260505H and 260505K; Part No. 00055430010 UDI-DI 08426950807759 Lots 260505H, 260505K and 260505F; Part No. 00055430011 UDI-DI 08426950807766 Lot 260505K; Product No. 00055445010 UDI-DI 08426950807810 Lots 260504AA and 260505K; Product No. 00055445011 UDI-DI 08426950807827 Lots 260505F and 260505J; Product No. 00055360010.A UDI-DI 08426950807599 Lot 260504AF;
Why
The active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
Distributed
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
Company
Instrumentation Laboratory
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2900-2026 · retrieved 6 August 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.