TemporaryFDAOngoing
Can cause harm that is usually temporary.
GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.
Check the device or its packaging for
Part No.00077407508 UDI-DI 00195226000764 Lot 260504E; Part No. 00077407510 UDI-DI 00195226000665 Lot 260505B; Part No. 00077415004 UDI-DI 00195226000856 Lot 260505A; Part No. 00077415008 UDI-DI 00195226000757 Lot No.260505A; Part No. 00077415010 UDI-DI 00195226000658 Lot 260505C; Part No. 00077430010 UDI-DI 00195226000641 Lot 260505B
- Why
- The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.
- Distributed
- Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.
- Company
- Instrumentation Laboratory
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-2899-2026 · retrieved 6 August 2026