Atellica CH Urine Albumin REF: 11537225, in vitro diagnostic

Check the device or its packaging for

All Lots within expiry/UDI: 00630414611099
Why
Processing of urine albumin assay on the CI analyzers may lead to the following issues: Falsely depressed Atellica CH Urea Nitrogen (UN_c) results; Falsely depressed Atellica CH Enzymatic Hemoglobin A1c (A1c_E) results; Increased rate of reaction cuvette QC failures and Delayed results due to QC failures.
Distributed
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Austria, Bahamas, Belgium, Brazil, Canada, Chile, Curacao, St. Eus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel,¿Italy,¿Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Republic of Korea, Romania, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.¿