TemporaryFDAOngoing

Can cause harm that is usually temporary.

Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900

Check the device or its packaging for

Flow-i C20 6677200 / UDI-DI 07325710001349 Flow-i C30 6677300 / UDI-DI 07325710001363 Flow-i C40 6677400 / UDI-DI 07325710001318 Flow-i C20 6888520 / UDI-DI 07325710010617 Flow-i C30 6888530 / UDI-DI 07325710010624 Flow-i C40 6888540 / UDI-DI 07325710010631 Flow-c 6887700 / UDI-DI 07325710009765 Flow-e 6887900 / UDI-DI 07325710010457 Software versions 4.8, 4.10, and 4.11.
Why
Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
Distributed
Worldwide - US Nationwide distribution to the states of AL, AR, CA, CO, DC, DE, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, SC, TX, VA, WA, WV.
Company
Maquet Critical Care AB
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2860-2026 · retrieved 3 August 2026