Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Brand Name: Quantum Rehab Product Name: Q6 Edge 3 Model/Catalog Number: Q6 Edge 3 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

Check the device or its packaging for

Lot Code: Q6 Edge 3 0060650920015*8 K143383 D246699 41875 ITI
Why
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Distributed
Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.
Company
Pride Mobility Products Corp.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2849-2026 · retrieved 31 July 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.

Local Recall