Can cause harm that is usually temporary.
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system
Check the device or its packaging for
GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Argentina: GEM Premier 5000 Serial Nos.: 25100466, 24097933 and 23064897. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- Why
- Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
- Distributed
- US Nationwide distribution.