TemporaryFDAOngoing
Can make you unwell for a while. Do not eat it.
EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222
Check your package for
UDI-DI: 20763000135700, Lot Numbers: 0232602300 0232602383 0232612916 0232680078 0232708967 0232709001 0232910880 0232910881 0232910882 0233077983 0233245850 0233245860 0233245980; UDI-DI: 00763000135706, Lot Numbers: 0232612916 0232709001
- Why
- Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
- Distributed
- Worldwide - US Nationwide distribution.
- Company
- Medtronic Perfusion Systems
- Announced
- Recall began