Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622

Check the device or its packaging for

UDI-DI: 20763000135670, Lot Numbers: 0232601501 0232601510 0232601549 0232601550 0232602178 0232602290 0232602302 0232910952 0232910953 0232910954 0232910955 0233244075 0233244301 0233244531 0233286362 0233286374 0233286390 0233409034 0233416071 0233528731
Why
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Distributed
Worldwide - US Nationwide distribution.
Company
Medtronic Perfusion Systems
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2808-2026 · retrieved 27 July 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.