TemporaryFDAOngoing
Can cause harm that is usually temporary.
Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3
Check the device or its packaging for
All Units operating with firmware version 1.26.2/UDI:(01)17330818345674
- Why
- This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
- Distributed
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CAN.
- Company
- Permobil Inc
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-2795-2026 · retrieved 24 July 2026