TemporaryFDAOngoing
Can cause harm that is usually temporary.
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116G
Check the device or its packaging for
DYKMBNDL116G UDI-DI 10198459156991 Lots 25GBG086 25JBF391 25KBA114 25KBN017 25LBQ601
- Why
- Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
- Distributed
- US Nationwide distribution in the states of CA, FL, MD. PA, and TX.
- Company
- Medline Industries, LP
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-2787-2026 · retrieved 23 July 2026
Read the official notice