TemporaryFDAOngoing
Can make you unwell for a while. Do not eat it.
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
Check your package for
Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0
- Why
- For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
- Distributed
- Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.
- Company
- Medtronic MiniMed, Inc.
- Announced
- Recall began