Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.

Check the device or its packaging for

Software Version - 2.0.0 GS1 GTIN code: 199874903484 UDI code: GUDID registration is in progress
Why
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
Distributed
U.S. Nationwide distribution in the states of AR, AZ, CA, CO, FL. IA, ID, NC, NJ, PA, TX and WI.
Company
Emano Metrics, Inc.
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2777-2026 · retrieved 22 July 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.