FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21M402, IFU Reference Number LBL1062
Check your package for
UDI-DI: 00810087180072; Lots: CL12697, CL12698, CL12701, CL12702, CL12824, CL12832, CL12833, CL12861, CL12888, CL12889, CL12912, CL12913, CL12916, CL12918, CL12968, CL12975, CL12977, CL13003, CL13016, CL13019, CL13045, CL13047, CL13059, CL13078, CL13079, CL13101, CL13102, CL13120, CL13121, CL13140, CL13141, CL13158, CL13159, CL13195, CL13211, CL13218, CL13232, CL13233, CL13259, CL13290, CL13339, CL13367, CL13400, CL13421, CL13516, CL13518, CL13572, CL13577, CL13615, CL13645, CL13669, CL13707, CL13720, CL13733, CL13821, CL13834, CL13857, CL13926, CL13941, CL13976, CL13989, CL14026, CL14052, CL14078, CL14121, CL14138, CL14170, CL14186, CL14225, CL14248, CL14285, CL14344, CL14352, CL14374, CL14395, CL14438, CL14487, CL14499, CL14517, CL14580, CL14632, CL14669, CL14690, CL14710, CL14748, CL14768, CL14967, CL15004, CL15008, CL15066, CL15106, CL15127, CL15190, CL15238, CL15271, CL15317, CL15357, CL15396, CL15426, CL15473, CL15489, CL15567, CL15636, CL15639, CL15753, CL15774, CL15807
- Why
- Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
- Distributed
- Global Distribution. US Nationwide.
- Company
- Boston Scientific Corporation
- Announced
- Recall began