Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Check your package for

Serial Numbers: 4-01440NB10CL, 2 4-01441NB10CL.
Why
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Distributed
US Nationwide distribution.
Company
Daavlin Distributing Company
Announced
Recall began