TemporaryFDAOngoing
Can cause harm that is usually temporary.
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Check the device or its packaging for
Serial Numbers: 4-01440NB10CL, 2 4-01441NB10CL.
- Why
- Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Distributed
- US Nationwide distribution.
- Company
- Daavlin Distributing Company
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-2742-2026 · retrieved 15 July 2026
Read the official notice