TemporaryFDAOngoing
Can cause harm that is usually temporary.
Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188.
Check the device or its packaging for
Model Number: 188. UDI-DI: 00821925001053. All Lot Numbers.
- Why
- Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
- Distributed
- Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.
- Company
- Olympus Corporation of the Americas
- Announced
- This is the date the recall was announced. A company may have started it weeks earlier.
- Recall began
Source: FDA record Z-2718-2026 · retrieved 10 July 2026
Read the official notice