Radius VSM Disposable NIBP Cuff, REF: 4825, 4826
Check the device or its packaging for
REF/UDI-DI/Lot: 4825/00843997006989/21G6S, 22JHV, 23H03, 23J05, 24DBV, 24NGG, 25ADG, E21D31, E21JWM, E23FVD; 4826/10843997006993/21HVX, 22ELB, 23G31, 23H04, 23HWL, 23J06, 24DBW, 24MNP, 24NGF, 25ADF, E21D32, E21F04, E21JFA, E21KHK
- Why
- Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
- Distributed
- Worldwide - US Nationwide distribution in the states of OK, CA , MD , MA , NY , UT , NC , TN , MN , TX , OH , PA , VA and the countries of Italy, Latvia, Iraq, Saudi Arabia, Germany, Spain, Turkey, France, Australia, Malta, Oman, United Arab Emirates, Martinique, Netherlands, Cyprus, Kuwait, Czech Republic, United Kingdom, Croatia, Switzerland, Belgium, Singapore, Romania, Colombia, Hong Kong, Israel, Finland, Poland, New Zealand, Tunisia, Chile, Algeria, Canada, Denmark, Palestinian Territory, Qatar, Greece, Paraguay.