Español
TemporaryFDAOngoing

Can cause harm that is usually temporary.

Radius VSM Disposable NIBP Cuff, REF: 4825, 4826

Check the device or its packaging for

REF/UDI-DI/Lot: 4825/00843997006989/21G6S, 22JHV, 23H03, 23J05, 24DBV, 24NGG, 25ADG, E21D31, E21JWM, E23FVD; 4826/10843997006993/21HVX, 22ELB, 23G31, 23H04, 23HWL, 23J06, 24DBW, 24MNP, 24NGF, 25ADF, E21D32, E21F04, E21JFA, E21KHK
Why
Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
Distributed
Worldwide - US Nationwide distribution in the states of OK, CA , MD , MA , NY , UT , NC , TN , MN , TX , OH , PA , VA and the countries of Italy, Latvia, Iraq, Saudi Arabia, Germany, Spain, Turkey, France, Australia, Malta, Oman, United Arab Emirates, Martinique, Netherlands, Cyprus, Kuwait, Czech Republic, United Kingdom, Croatia, Switzerland, Belgium, Singapore, Romania, Colombia, Hong Kong, Israel, Finland, Poland, New Zealand, Tunisia, Chile, Algeria, Canada, Denmark, Palestinian Territory, Qatar, Greece, Paraguay.
Company
Masimo Corporation
Announced
This is the date the recall was announced. A company may have started it weeks earlier.
Recall began

Source: FDA record Z-2671-2026 · retrieved 2 July 2026

Read the official notice

Every recall is shown word for word from the FDA and USDA, with a link to the source.