Español
TemporaryFDAOngoing

Can make you unwell for a while. Do not eat it.

Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.

Check your package for

Model Number: 867232. UDI-DI: (01)00884838102842. Software Version Number: 4.0.0.0. Serial Numbers: US001A002F, US001A00AW, US001A00C9, US001A00EW, US001A00F0, US001A00GM, US001A00N4, US001A00PB, US001A00PZ, US001A00R1, US001A00UE, US001A00VU, US001A00Z4, US001A013T, US001A018A, US001A018R.
Why
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
Distributed
US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.
Company
Philips North America Llc
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.