Octopus Evolution AS Tissue Stabilizer, Model TS2500

Check your package for

UDI-DI: 00643169668904, 00613994764379; Serial Numbers: 0232670681 0232709164 0232709168 0232709194 0232709194 0232709199
Why
During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Distributed
Worldwide - US Nationwide distribution.
Company
Medtronic Perfusion Systems