Español
SeriousFDAOngoing

Can cause serious illness or death. Do not eat it.

Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device

Check your package for

UDI: 10887761985015; Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Why
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Distributed
Worldwide distribution - US Nationwide and the country of Canada.
Company
Baxter Healthcare Corporation
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.