Español
TemporaryFDAOngoing

Can make you unwell for a while. Do not eat it.

iLet Bionic Pancreas, REF: BB1001

Check your package for

iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)
Why
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
Distributed
US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL
Company
Beta Bionics, Inc.
Announced
Recall began
Read the official notice

Every recall — food, medication, or medical device — is shown word-for-word from the official FDA or USDA notice, with a link straight to it. Nothing here is summarised, reworded, or filled in. This site is not affiliated with any government agency.